3526 articles – 5250 references  [version française]

inria-00321688, version 1

A methodology for validating a 3D imaging modality for brain AVM delineation: Application to 3DRA.

Marie-Odile Berger () 1, René Anxionnat 123, Erwan Kerrien (Author to contact preferably) 1, Luc Picard 23, Michael Söderman a4

Computerized Medical Imaging and Graphics 32, 7 (2008) 544-553

Abstract: A general methodology is described to validate a 3D imaging modality with respect to 2D digital subtracted angiography (DSA) for brain AVMs (BAVM) delineation. It relies on the assessment of the statistical compatibility of the radiosurgical target delineated in 3D with its delineations in 2D. This methodology is demonstrated through a preliminary evaluation of rotational 3D angiography (3DRA). Generally speaking, BAVM delineation cannot be performed on 3DRA alone. However, in our study, 3DRA showed similar performances to DSA for rather easy cases, and even better for 3 patients. Conversely, 3 problematic cases are identified and discussed.

  • a –  Institution for clinical neurosciences - section of neuroradiology
  • 1:  MAGRIT (INRIA Lorraine - LORIA)
  • CNRS : UMR7503 – INRIA – Université Henri Poincaré - Nancy I – Université Nancy II – Institut National Polytechnique de Lorraine (INPL)
  • 2:  Département de neuroradiologie diagnostique et thérapeutique
  • CHU Nancy – Université Henri Poincaré - Nancy I
  • 3:  UHP
  • Université Henri Poincaré - Nancy I
  • 4:  Karolinska Institutet
  • Karolinska Institutet
  • Domain : Computer Science/Computer Vision and Pattern Recognition
    Life Sciences/Human health and pathology
  • Keywords : Angiography – Brain Arteriovenous Malformation – Radiosurgery – Validation – Principal Component Analysis
 
  • inria-00321688, version 1
  • oai:hal.inria.fr:inria-00321688
  • From: 
  • Submitted on: Wednesday, 5 November 2008 10:18:53
  • Updated on: Wednesday, 5 November 2008 10:19:58