Model of a Vascular C-Arm for 3D Augmented Fluoroscopy in Interventional Radiology

Sébastien Gorges () 12, Erwan Kerrien 2, Marie-Odile Berger () 2, Yves Trousset 1, Jérémie Pescatore 1, René Anxionnat 23, Luc Picard 3

8th International Conference on Medical Image Computing and Computer Assisted Intervention- MICCAI'05 3750 (2005) 214-222

Résumé : This paper deals with the modeling of a vascular C-arm to generate 3D augmented fuoroscopic images in an interventional radiology context. A methodology based on the use of a multi-image calibration is proposed to assess the physical behavior of the C-arm. From the knowledge of the main characteristics of the C-arm, realistic models of the acquisition geometry are proposed. Their accuracy was evaluated and experiments showed that the C-arm geometry can be predicted with a mean 2D reprojection error of 0.5 mm. The interest of 3D augmented uoroscopy is also assessed on a clinical case.

  • 1 :  General Electric Medical Systems [Buc] (GE Healthcare)
  • General Electric Medical Systems
  • 2 :  MAGRIT (INRIA Lorraine - LORIA)
  • CNRS : UMR7503 – INRIA – Université Henri Poincaré - Nancy I – Université Nancy II – Institut National Polytechnique de Lorraine (INPL)
  • 3 :  Département de neuroradiologie diagnostique et thérapeutique
  • CHU Nancy – Université Henri Poincaré - Nancy I
  • Domaine : Informatique/Autre
 
  • inria-00000386, version 1
  • oai:hal.inria.fr:inria-00000386
  • Contributeur : 
  • Soumis le : Vendredi 4 Novembre 2005, 00:10:01
  • Dernière modification le : Jeudi 26 Juillet 2007, 11:19:27